Beckman-Coulter FC-500 Analyzer
   
   

The Beckman-Coulter FC-500 instrument has a 488nm argon laser and a 637nm HeNe laser and can simultaneously measure up to 5 color parameters.

Parameter Spectra Fluorochrome
FLS 488nm N/A
SS 488nm N/A
FL1 525nm FITC , GFP, YFP, CFSE,
FL2 575nm PE, DsRed, PI
FL3 610nm PETxRd, PI, Alexa610
FL4 675nm PECy5, APC, 7AAD, PerCP
FL5 750nm PECy7, APC-Cy7

Note: APC and APC-Cy7 are excited by the 637nm HeNe Laser.